Neonatal/Pediatric Deltran

Neonatal/Pediatric Deltran® Plus is a closed blood collection system that is designed specifically to conserve blood and protect the neonatal patient from infection.

Product Description
Models
Specifications
Deltran® Brochure

Deltran Neonatal/Pediatric Blood Pressure Monitoring and Blood Sampling System.The standards for accuracy, patient safety and ease of use you get with Deltran® disposable blood pressure transducers and Deltran® Plus arterial blood collection systems are now available in neonatal/pediatric versions.

Closed system reduces the risk of infection.

All blood volume is returned back to the patient.

The plastic cannula with a rounded tip provides superior clinician protection from needle stick.

The integrated 3cc reservoir syringe eliminates the splashing and spraying of blood during sampling.

One-handed operation of the reservoir syringe makes the system easy to use.

These products reflect the more than 25 years of experience of Utah Medical Products® in critical care monitoring, and the company's commitment to the constant refinement of technology in pursuit of the highest value for clinician and patient.

Neonatal/Pediatric Deltran® Blood Pressure Transducers
Model Number Description Unit of Measure Image
DPT-300 Deltran® Disposable Pressure Transducer, NEONATAL/PEDIATRIC with Integral
30cc Flush Device and Stopcock. *
25/BOX Deltran® Disposable Blood Pressure Transducer, NEONATAL/PEDIATRIC with Integral 30cc Flush Device and Stopcock. Model DPT-300
DPT-324 Deltran® Disposable Pressure Transducer System, NEONATAL/PEDIATRIC with Integral 30cc Flush Device and Stopcock - Includes: Deltran® DPT w/30cc Flush and Stopcock, 24" pressure tubing, and two 3-Way Stopcocks. * 25/BOX Deltran® Disposable Blood Pressure Transducer System, NEONATAL/PEDIATRIC with Integral 30cc Flush Device and Stopcock. Model DPT-324
 
Neonatal/Pediatric Needleless Blood Collection
Model Number Description Unit of Measure Image
ABC-328NP Deltran® Plus Needleless Arterial Blood Collection System, NEONATAL/PEDIATRIC with Integral 30cc Flush Device and Stopcock - Includes: Deltran® DPT w/30cc Flush and Stopcock, 6cc Reservoir Syringe, one ea. 2", 4" and 20" pressure tubing, one 1-Way Stopcock, two ea. 3-Way Stopcocks, and one Labsite Access Port. *
10/BOX Deltran® Plus Needleless Arterial Blood Collection System, NEONATAL/PEDIATRIC with Integral 30cc Flush Device and Stopcock. Model ABC-328NP
ABC-350NP Deltran® Plus Needleless Arterial Blood Collection System, NEONATAL/PEDIATRIC with Integral 30cc Flush Device and Stopcock - Includes: Deltran® DPT w/30cc Flush and Stopcock, 6cc Reservoir Syringe, 50" pressure tubing, one 1-Way Stopcock, and one Labsite Access Port. *
10/BOX Deltran® Plus Needleless Arterial Blood Collection System, NEONATAL/PEDIATRIC with Integral 30cc Flush Device and Stopcock. Model ABC-350NP
ABC-050NP NEONATAL/PEDIATRIC Needleless Arterial Blood Collection System W/O Transducer. Includes: 50" Pressure Tubing, 6cc Reservoir Syringe, one 1-Way Stopcock, and one Labsite Access Port. *
10/BOX NEONATAL/PEDIATRIC Needleless Arterial Blood Collection System W/O Transducer. Model ABC-050NP
 
Cannulas
Model Number Description Unit of Measure Image
ABC-100 Protective Safety Cannula for Deltran® Plus Systems. * 100/BOX Protective Safety Cannula for Deltran® Plus Systems. Model ABC-100
 

* Product is CE Marked

Deltran® Disposable Pressure Transducer
Continuous Flow Rate (Applies to models with integral flush device) 3cc/hr (+2/-1cc/hr) or 30cc/hr (± 10cc/hr) at 300 mmHg
Operating Pressure Range -50 to +300 mmHg
Sensitivity 5 µV/V/mmHg, ±2% (typically < ±1%)
Zero Drift With Time < ±1.0 mmHg/8 hours after 10 min. warm-up to operating temperature
Leakage Current <2 µA @ 115 Vac rms at 60 Hz
Unbalance ±75 mmHg
Overpressure Protection -400 to +4000 mmHg
Operating Temperature 15° C to 40° C
Excitation Voltage and Frequency 2 to 10 Vdc; or Vac rms to 5 kHz
Operating Life >500 hours
Storage Temperature -25° C to +65° C
Defibrillation Withstand 5 discharges/5 minutes of 400 joules @ 50 ohm load
Natural Frequency >200 Hz in saline
Phase Shift <5° at 5 kHz
Output Impedance 270 Ohms to 400 Ohms
Input Impedance 270 Ohms to 400 Ohms
 

 