Deltran® is a disposable pressure transducer that provides consistent and accurate readings during invasive blood pressure monitoring.
Product Description
Models
Specifications
Deltran®
Brochure ![]()
Deltran®
disposable blood pressure transducers reflect more than 25 years of
experience in critical care blood pressure monitoring and it is our
goal to continually seek ways to improve accuracy and safety, and
to make Deltran® blood pressure monitoring products the easiest
to use.
Consistent and accurate readings during blood pressure monitoring.
Components that simplify setup.
The transducer can easily be mounted on either the patient, bed or
IV pole.
Fully integrated flow-thru component design simplifies filling and
debubbling.
Versatile, dual function Snap-Tab™ allows dynamic response testing.
Superior fluid path visualization.
3cc or 30cc per hour flow rates.
With a variety of interface cables available (see Deltran®
Accessories), Deltran® blood pressure monitoring systems can
be connected to most types of monitors.
Since 1978, Utah Medical Products® has been
a leader in disposables for blood pressure monitoring. By continually
striving to anticipate the critical needs of clinicians, Utah Medical
is constantly refining technology to develop transducer products of
the highest value to clinicians and patients.
With the introduction of Deltran® disposable pressure transducers,
Utah Medical established new standards for patient safety and ease
of use, standards by which competitors are measured.
* Product is CE Marked
| Deltran® Disposable Pressure Transducer | |
| Continuous Flow Rate (for models with integral flush device) | 3cc/hr (± 1cc/hr) or 30cc/hr (± 10cc/hr) at 300 mmHg |
| Operating Pressure Range | -50 to +300 mmHg |
| Sensitivity | 5 µV/V/mmHg, ±2% (typically < ±1%) |
| Zero Drift With Time | < ±1.0 mmHg/8 hours after 10 min. warm-up to operating temperature |
| Leakage Current | <2 µA @ 115 Vac rms at 60 Hz |
| Unbalance | ±75 mmHg |
| Overpressure Protection | -400 to +4000 mmHg |
| Operating Temperature | 15° C to 40° C |
| Excitation Voltage and Frequency | 2 to 10 Vdc; or Vac rms to 5 kHz |
| Operating Life | >500 hours |
| Storage Temperature | -25° C to +65° C |
| Defibrillation Withstand | 5 discharges/5 minutes of 400 joules @ 50 ohm load |
| Natural Frequency | >200 Hz in saline |
| Phase Shift | <5° at 5 kHz |
| Output Impedance | 270 Ohms to 400 Ohms |
| Input Impedance | 270 Ohms to 400 Ohms |